Vulnerabilities

CVE-2025-24836

Medium 6.1 CVSS 4.0

Description

With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.

Description as published by the source, in English.

Key facts

Published
2025-02-13
Last modified
2026-06-17
Assigned by
ics-cert@hq.dhs.gov
CVSS vector
CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:
Scored by
ics-cert@hq.dhs.gov
Probability of exploitation
0.2 % higher than 15 % of all known vulnerabilities
EPSS, next 30 days, model v2026.06.15
Fix available
Not registered in the sources

Affected products

Vendor Product Versions Status
Qardio Heart Health Android Mobile Application 2.5.1 Affected
Qardio Heart Health IOS Mobile Application 2.7.4 Affected
Qardio QardioARM All versions Affected

Weakness type