CVE-2025-14096
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Description as published by the source, in English.
Key facts
- Published
- 2025-12-17
- Last modified
- 2026-06-17
- Assigned by
- 46b595e9-1acc-41cb-9398-adaf98d37a9b
- CVSS vector
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H- Scored by
- 46b595e9-1acc-41cb-9398-adaf98d37a9b
- Probability of exploitation
-
0.1 %
higher than 2 %
of all known vulnerabilities
EPSS, next 30 days, model v2026.06.15 - Fix available
- Not registered in the sources
Affected products
| Vendor | Product | Versions | Status |
|---|---|---|---|
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX analyzers (Windows 7) | Windows 7 Operating system | Affected |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX analyzers (Windows 7) | Windows XP Operating system | Affected |
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers (Windows XP) | Windows 7 Operating system | Affected |
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers (Windows XP) | Windows XP Operating system | Affected |
| Radiometer Medical Aps | AQT90 FLEX Analyzers (Windows 7) | Windows 7 Operating system | Affected |
| Radiometer Medical Aps | AQT90 FLEX Analyzers (Windows 7) | Windows XP Operating system | Affected |
Weakness type
-
CWE-250: Execution with Unnecessary Privileges
The product performs an operation at a privilege level that is higher than the minimum level required, which creates new weaknesses or amplifies the consequences of other weaknesses.
-
CWE-798: Use of Hard-coded Credentials
The product contains hard-coded credentials, such as a password or cryptographic key.
Attack techniques