Vulnerabilities

CVE-2025-1204

High 7.7 CVSS 4.0

Description

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

Description as published by the source, in English.

Key facts

Published
2025-02-25
Last modified
2026-06-17
Assigned by
ics-cert@hq.dhs.gov
CVSS vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:
Scored by
ics-cert@hq.dhs.gov
Probability of exploitation
0.5 % higher than 38 % of all known vulnerabilities
EPSS, next 30 days, model v2026.06.15
Fix available
Not registered in the sources

Affected products

Vendor Product Versions Status
Contec Health CMS8000 Patient Monitor 0 Affected

Weakness type

  • CWE-912: Hidden Functionality

    The product contains functionality that is not documented, not part of the specification, and not accessible through an interface or command sequence that is obvious to the product's users or administrators.

References

Does this affect you?

A vulnerability matters only if you run the product, in an affected version, somewhere an attacker can reach.

Check your asset list Browse all vulnerabilities