Vulnerabilities

CVE-2025-0683

High 8.2 CVSS 4.0

Description

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Description as published by the source, in English.

Key facts

Published
2025-01-30
Last modified
2026-06-17
Assigned by
ics-cert@hq.dhs.gov
CVSS vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:
Scored by
ics-cert@hq.dhs.gov
Probability of exploitation
0.8 % higher than 52 % of all known vulnerabilities
EPSS, next 30 days, model v2026.06.15
Fix available
Not registered in the sources

Affected products

Vendor Product Versions Status
Contec Health CMS8000 Patient Monitor All versions Affected

Weakness type

Attack techniques

Derived through the weakness type: CVE to CWE to CAPEC to ATT&CK. This shows which techniques the weakness class is historically used for. It does not assert that this vulnerability has been exploited that way.

  • T1606.001: Web Cookies credential-access

References

Does this affect you?

A vulnerability matters only if you run the product, in an affected version, somewhere an attacker can reach.

Check your asset list Browse all vulnerabilities