CVE-2025-0626

High 7.7 CVSS 4.0

Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Description as published by the source, in English.

Key facts

Published
2025-01-30
Last modified
2026-06-17
Assigned by
ics-cert@hq.dhs.gov
CVSS vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:
Scored by
ics-cert@hq.dhs.gov
Probability of exploitation
1.1 % higher than 63 % of all known vulnerabilities
EPSS, next 30 days, model v2026.06.15
Fix available
Not registered in the sources

Affected products

Vendor Product Versions Status
Contec Health CMS8000 Patient Monitor All versions Affected

Weakness type

  • CWE-912: Hidden Functionality

    The product contains functionality that is not documented, not part of the specification, and not accessible through an interface or command sequence that is obvious to the product's users or administrators.