Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Description as published by the source, in English.

Key facts

Published
2024-11-14
Last modified
2026-06-17
Assigned by
productsecurity@baxter.com
CVSS vector
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Scored by
productsecurity@baxter.com
Probability of exploitation
0.2 % higher than 7 % of all known vulnerabilities
EPSS, next 30 days, model v2026.06.15
Fix available
Not registered in the sources

Affected products

Vendor Product Versions Status
Baxter Life2000 Ventilation System 06.08.00.00 and prior Affected

Weakness type

Attack techniques

Derived through the weakness type: CVE to CWE to CAPEC to ATT&CK. This shows which techniques the weakness class is historically used for. It does not assert that this vulnerability has been exploited that way.

  • T1078: Valid Accounts stealth, persistence, privilege-escalation, initial-access
  • T1110.001: Password Guessing credential-access
  • T1110.003: Password Spraying credential-access
  • T1110.004: Credential Stuffing credential-access
  • T1558: Steal or Forge Kerberos Tickets credential-access