Vulnerabilities

CVE-2024-48973

Critical 9.3 CVSS 3.1

Description

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Description as published by the source, in English.

Key facts

Published
2024-11-14
Last modified
2026-06-17
Assigned by
productsecurity@baxter.com
CVSS vector
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Scored by
productsecurity@baxter.com
Probability of exploitation
0.2 % higher than 13 % of all known vulnerabilities
EPSS, next 30 days, model v2026.06.15
Fix available
Not registered in the sources

Affected products

Vendor Product Versions Status
Baxter Life2000 Ventilation System 06.08.00.00 and prior Affected

Weakness type

  • CWE-1263: Improper Physical Access Control

    The product is designed with access restricted to certain information, but it does not sufficiently protect against an unauthorized actor with physical access to these areas.

References

Does this affect you?

A vulnerability matters only if you run the product, in an affected version, somewhere an attacker can reach.

Check your asset list Browse all vulnerabilities