CVE-2025-0626

Høy 7.7 CVSS 4.0

Beskrivelse

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Beskrivelsen gjengis slik kilden har publisert den, på engelsk.

Nøkkelfakta

Publisert
2025-01-30
Sist endret
2026-06-17
Tildelt av
ics-cert@hq.dhs.gov
CVSS-vektor
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:
Scoret av
ics-cert@hq.dhs.gov
Sannsynlighet for utnyttelse
1.1 % høyere enn 63 % av alle kjente sårbarheter
EPSS, neste 30 dager, modell v2026.06.15
Rettelse finnes
Ikke registrert hos kildene

Berørte produkter

Leverandør Produkt Versjoner Status
Contec Health CMS8000 Patient Monitor All versions Berørt

Svakhetstype

  • CWE-912: Hidden Functionality

    The product contains functionality that is not documented, not part of the specification, and not accessible through an interface or command sequence that is obvious to the product's users or administrators.